Operation Steps:
Turn on the power switch.
If the door is not open, press [OPEN DOOR] to open it.
Check that the rotor is properly installed.
Place samples symmetrically and horizontally in the rotor.
Close the door (ensure you hear a “click”).
Enter the operation parameters:
[ROTOR]: Enter the rotor number; all rotor numbers are pre-programmed. When pressed, the number will blink—use the up/down keys to adjust, then press ENTER.
[RPM]: Set the speed in revolutions per minute; when pressed, the number will blink—adjust with up/down keys, then press ENTER.
[RCF]: Set the centrifugal force in g; when pressed, the number will blink—adjust with up/down keys, then press ENTER.
[TIME]: Set the duration; when pressed, the number will blink—adjust with up/down keys, then press ENTER.
[TEMP]: Set the temperature; when pressed, the number will blink—adjust with up/down keys, then press ENTER.
[ACCEL]: Acceleration setting; 0–9 levels available, higher number = faster acceleration.
[DECEL]: Deceleration setting; 0–9 levels available, higher number = faster deceleration.
[ENTER]: Confirm key; press this after setting any parameter.
Press [START] to begin operation. The centrifuge will automatically stop when the time reaches zero.
To stop midway, press [STOP] or [FAST STOP].
After operation stops, press [OPEN DOOR] to retrieve the samples.
Maintenance:
Keep the chamber clean.
Keep the rotor clean.
Regularly clean and lubricate the door latch and shaft (use Spinkote).
Precautions:
Do not open the lid or move the machine during operation.
PULSE: Short-run operation; hold this button to accelerate to the set speed, release to stop as quickly as possible.
Pre-cool the machine when operating at low temperatures.
Do not centrifuge hazardous materials (toxic, radioactive, pathogenic, etc.) without proper safety handling to avoid contamination or accidents due to tube breakage.